This FID peak normalization calculator calculates amount % for up to six compounds using peak area and relative response factors (RRF). Results can be normalized to 100% or a user-defined total, making it suitable for ASTM methods such as D6729 and D6730.
FID Peak Normalization Calculator Interface
What inputs are required ?
- Peak area and Relative Response Factor of compound 1 to compound 6
- Optional: Referenced Amount% of compound 1 to compound 6
- Total normalization value in percentage
- If the required amount% is by mole, the relative response factor must also be by mole. Likewise for mass and volume
- Use FID Relative Response Factor Calculator for GAS to find the relative response factor when the sample is in gas phase
- Use FID Relative Response Factor Calculator for LIQUID to find the relative response factor when the sample is in liquid phase
What the FID Peak Normalization Calculator Outputs
- Amount% of compound 1 to compound 6
- Relative Difference Percentage between Referenced Amount% and Calculated Amount% of compound 1 to compound 6
- The unit of amount% will be by mole if the provided RRF is by mole. Likewise for mass and volume.
Before You Go
Want to understand how this calculator works behind the scenes?
Read the full explanation in this Flame Ionization Detector (FID) series in the GC Hardware.
Browse related calculators below:
- FID RRF Calculator for Unknown GC Peaks
- FID RRF Verification Calculator (Theoretical vs Experimental)
- FID Methanizer RRF Calculator
Or if you’d like to explore other free GC calculators, browse the full collection here.